Staircase Galvanostatic

In staircase galvanostatic measurements, a current scan is performed using discrete current steps with a constant step height.

The potential and current are measured over time at a defined sampling rate.

The galvanostatic staircase can be terminated early if a defined potential limit Emin/Emax is exceeded.

Parameter Description

Parameter

Name

Description

Unit

I start

start current

starting DC current

A

I step

step current

current of each step for the staircase function

A

I end

end current

maximum DC current of the last step

A

t settle

settle time

settling time at I start before the first V-I step is stimulated

s

t step

step duration

duration of each step of the staircase function of V-I

s

ODR

output data rate

number of measurement points per second
note: the current is always measured at the maximum possible speed and integrated for the selected data rate

\(\frac{1}{s}\)

E min

min potential

minimum potential limit for premature staircase determination

V

E max

max potential

maximum potential limit for premature staircase determination

V

A Start Phase Galvanostatic can be enabled or disabled before the method is executed.

../../_images/staircase_galvanostatic_desc.svg

Measurement Result

../../_images/staircase_galvanostatic_result.svg

The voltage and current curves are recorded as a function of time. Additional measurement signals are also recorded when activated.

Custom Experiment Builder

This experiment is a combination of the following blocks: