Staircase Galvanostatic
In staircase galvanostatic measurements, a current scan is performed using discrete current steps with a constant step height.
The potential and current are measured over time at a defined sampling rate.
The galvanostatic staircase can be terminated early if a defined potential limit Emin/Emax is exceeded.
Parameter Description
Parameter |
Name |
Description |
Unit |
|---|---|---|---|
I start |
start current |
starting DC current |
A |
I step |
step current |
current of each step for the staircase function |
A |
I end |
end current |
maximum DC current of the last step |
A |
t settle |
settle time |
settling time at I start before the first V-I step is stimulated |
s |
t step |
step duration |
duration of each step of the staircase function of V-I |
s |
ODR |
output data rate |
number of measurement points per second |
\(\frac{1}{s}\) |
E min |
min potential |
minimum potential limit for premature staircase determination |
V |
E max |
max potential |
maximum potential limit for premature staircase determination |
V |
A Start Phase Galvanostatic can be enabled or disabled before the method is executed.
Measurement Result
The voltage and current curves are recorded as a function of time. Additional measurement signals are also recorded when activated.