Staircase Voltammetry

In staircase voltammetry, a potential scan is performed using discrete potential steps with a constant step height.

The potential and current are measured over time at a defined sampling rate.

The potentiostatic staircase can be terminated early if a defined current limit Imin/Imax is exceeded.

Parameter Description

Parameter

Name

Description

Unit

E start

start potential

starting DC potential
can be set in absolute potential or relative to OCV

V

E step

step potential

potential of each step for the staircase function

V

E end

end potential

maximum DC potential of the last step
can be set in absolute potential or relative to OCV

V

t settle

settle time

settling time at E start before the first I-V step is stimulated

s

t step

step duration

duration of each step of the staircase function of I-V

s

ODR

output data rate

number of measurement points per second
autoranging is only available for rates up to 50 Hz
note: the current is always measured at the maximum possible speed and integrated for the selected data rate

\(\frac{1}{s}\)

I range

current range (abs)

optional fixed current range selection if autoranging is disabled

A

I min

minimum current

minimum current limit for premature staircase determination

A

I max

maximum current

maximum current limit for premature staircase determination

A

A Start Phase Potentiostatic can be enabled or disabled before the method is executed.

Diagram of the staircase voltammetry measurement, illustrating the discrete potential steps of constant step height applied over time.

Measurement Result

Example result of a staircase voltammetry experiment, showing the voltage and current curves recorded as a function of time.

The voltage and current curves are recorded as a function of time. Additional measurement signals are also recorded when activated.